The Durability Testing of Porous Bifurcated Stent/Grafts Without Restricting Leakage
James C. Conti,1,2 Elaine R Strope,1 Donald J. Rohde,1 Richard Carp,1
Steven Kulan,3
1 Dynatek Dalta Scientific Instruments, Fourth and Main Sts., Galena, MO 65656 USA
2 Southwest Missouri State University, Dept. of Physics and Astronomy, Springfield, MO USA
3 Springfield Applied Mathmethics
Introduction:
The durability or fatigue testing
of porous vascular grafts or stent/grafts has, along
with the standard problems associated with
accelerated testing, the obvious additional
disadvantage that the testing fluids are
constantly leaked through the wall, thereby
reducing the volume of testing fluid available. In
the past the most common approach to this
problem was to use a material to restrict the
leakage of the fluid out of the sample during
testing. This was commonly done with either a
thin walled latex insert or by the utilization of
some blocking material such as gelatin to
accomplish the same restriction of flow. In some
instances, however, the utilization of these
materials causes an additional concern regarding
abrasion or modification of the three-dmensional
mechanical properties of the graft.
Purpose:
To evaluate the durability of a porous
bifurcated stent/graft, without the use of any
materials to restrict leakage.
Methods:
Initially the leakage of the stent/graft
designed for use in the aortic iliac bifurcated
position was determined using a pulsatile system
loading the graft at various pressures up to
150/110mm Hg. A device was then designed to
allow for the cyclic pressurization of these
bifurcated devices up to this pressure differential
at speeds up to 1200 cycles per minute (Figure 1).
This device included the utilization of a testing
fluid replacement circuit which was designed to
include a gravity feed and an impeller pump return loop. The key to the success of the device was the design of a large diameter quasi bileaflet silicone flapper valve that allowed for the replacement of the high volumes of fluid necessary to maintain the testing parameters at all cycle rates.
Results:
Leakage rates through the various stent/graft prototypes were as high as 4 liters per minute when tested at 150/110mm Hg at a pulsation rate of 72 beats per minute. It was found that monitoring the distal pressures at the exit points of the iliac artery portions of the stent/graft gave the most reliable data with the respect to assuring that the pulsed stent/graft was being cyclically loaded at pressures defined in the protocols.
Comments:
The device and protocols are currently being used to evaluate the various fatigue/durability/abrasion characteristics of a Nitinol Dacron bifurcated stent/graft. Details of the specific designs will be presented.
