High Frequency Testing of Nitinol Stent Material
25th Annual Meeting of the Society for Biomaterials, Transactions, 398, (1999)
James C. Conti
1, Elaine R. Strope, Ray Gregory
2
Dynatek Dalta Scientific Instruments, Fourth and Main St., Galena, MO 65656
1 Southwest Missouri State University, Dept. of Physics
2 Springfield Applied Mathematics
Introduction:
The use of Nitinol in implantable medical products has been increasing at a rapid
rate over the last several years. There are two reasons for this: The shape memory properties and
the superelasticity which make it highly convenient for the deployment of several medical products,
including stents and heart valves. Because of the need for reliable and well-designed
durability/fatigue experiments on these products, frequency response determinations of the
materials are important so that appropriate protocols can be used.
In most cases the shape memory properties of the Nitinol are trained into it after the fabrication of
the final product. As a result, it would be highly beneficial to develop raw material tests for
incoming material that are predictory of the performance and properties of the materials in their
final state. It is the purpose of this study to carry out initial investigations into the engineering
properties of Nitinol wire that predict the performance of the material when fabricated into the final
devised shape, as well as the speeds with which to test these final products.
Methods I:
Samples of Nitinol wire of varying diameters were subjected to four point bend
testing at a frequency of 1Hz. After that, the largest wire was subjected to four point bend testing
at 1, 20, and 40Hz, and three point bend testing at 1, 10, 20, and 40Hz. The testing was carried
out at room temperature and in air with an electromagnetically actuated micromechanical tester.
Results I:
Table 1 is a summary of the results obtained from the three point bend test. The
corrected simple modulus represents the slope of the stress-strain curve corrected for the cross sectional
area of the wire. This simplifies the calculations and allows for the observation of the
comparative properties of the material at different frequencies, rather than the absolute properties.
The hysteresis column represents the difference in the integrated area under the forward portion
minus the return portion of the cyclic loading curve.
As can be seen from the data, the apparent modulus of the material increases measurably when the
speed of the test changed from 20 to 40Hz. Interestingly, the largest change in hysteresis occurred
from 1Hz to 10Hz, Also, the increase in corrected simple modulus, when increasing the speed
from 20Hz to 40Hz, is not accompanied by a measurable change in the hysteresis of the test.
Table 2 is a summary of the four point bend testing. Again, this is a simple modulus test which
reflects the deflection of the contact points, rather than actual deflection of the bottom part of the
wire. As before, this was done in order to simplify the analysis, allowing comparison at the
different frequencies. In this instance, a large increase in corrected simple modulus is seen with
increase in wire diameter for testing at 1Hz.
These changes could, however, be a reflection of the fact that we are not measuring the deflection
of the lowest point in the wire. Also, the frequency change from 1Hz through 40Hz is
accompanied by an increase in the simple modulus and an increase in the percent of the input
energy lost through hysteresis.
Methods II:
Samples of Nitinol wires incorporated into actual stent/grafts were subjected to
three point bend testing at point spacings of 1 inch and 2 inches. Compression was delivered at a
deflection of 0.1 or 0.2mm at frequencies up to 100 Hz. Similar samples were subjected to four
point bend testing.
Results II:
Tables 3-7 show the results of these experiments. Similar to the first series, a
change in the simple modulus occurred at frequencies above 20 Hz. Figures 1-7 show how
increasing the frequency of the testing results just in increases in hysteresis and then erratic traces.
These erratic results could be the result of harmonics in the material.
Discussion:
Our attempt in this study was to compare three point and four point bend testing of
incoming raw material Nitinol wire to be used eventually for the fabrication of stents or other
implantable cardiovascular products, as well as Nitinol wire in the form of a stent. It was our
intention to evaluate how the apparent stress-strain relationships varied when testing at different
frequencies. Our loading was chosen to approximately replicate the type of loading that one might
expect in extreme conditions for a stent implantation, or individual barb loading on a stent graft.
The data seem to suggest that the properties of these materials change measurably at testing
frequencies above approximately 20Hz. It is apparent that a modification of the four point bend
test displacement measuring might be appropriate when generating the modulus numbers.
However, it is unlikely that the percentage hysteresis will change significantly in these tests with
that modification of displacement measuring.
The results indicate the need to be particularly careful when testing fatigue or durability properties
of incoming Nitinol material and final Nitinol products at frequencies above approximately 20Hz.
It is possible that modification of testing parameters might allow more confidence in high speed
testing; however, the complexity of Nitinol materials seem to render it somewhat different than
standard metal wire testing.