AAMI 24th Annual Meeting
James C. Conti, Ph.D.
Presentation:
Thank you Dr. Gabbay, Dr. Mueller, and thank you for
inviting me to this gathering. I am excited and hope that
this opens up a new wave of interest in in-vitro durability
testing.
The usefulness of experimental data is of course dependent
upon the reliability of that information, and any scientist
or engineer who would be queried concerning this fact would
be quick to agree, I am sure. Nonetheless, I am continually
confronted with attitudes from my colleagues concerning the
value of in-vitro heart valve durability testing, this from
individuals in laboratories that apparently feel that
attention to experimental detail is simply not important in
this particular case. At our facilities, we are continually
asked to write, review, or carry out protocols to evaluate
experimental heart valves. Although I don't think it is
appropriate to use this kind of a format for advertising
instruments, and I don't intent to do that, I would like to
quickly review the steps that we use to effectively
accomplish heart valve durability testing. As you will note
in the abstract there are certain assumptions that have been
made. The assumptions are basically that you have a heart
valve durability tester that does the job as you want it to
do. It is designed such that, as we have noted in the
abstract, the pressure measurements are free of inertial
interference, the transducers that you use are not frequency
dependent in the range in which you are using them, the
system has near zero capacitance, and the pressure
measurements do not change with time. I think that the
bottom line is that the pressure measurements that you make
must truly reflect the load that the valve is experiencing.
All those assumptions made, there is a series of
considerations that we go through when we set up to test a
heart valve. Three of these criteria are the same regardless
of whether it is a mechanical or a leaflet heart valve. The
first is that the valve must open fully each cycle, and it
must open smoothly. If it is a leaflet valve, the leaflet
cannot flutter as they are opening. If it is a mechanical
valve, the occluder cannot flutter as it is opening. After
full opening, a leaflet valve, be it polymeric or
bioprosthetic, must not bend over and hit the stents. It
should maintain a fairly parallel position with respect to
flow. On closing, again, the leaflets and the occluders
should not vibrate during that closing motion. At the end of
each cycle, the valve must experience full closure, and, in
particular with mechanical valves, right at the point of
closure one has to be careful to not experience any
cavitation. It is very difficult to always assess whether or
not cavitation is experienced, but if it is bad enough one
can notice what appears to be a flash or a mirror effect
flashing across the edge of the valve. Most of us understand
these different criteria for setting up a heart valve,
although when the time comes, most individuals that I work
with are usually under time constraints such that they have
a certain number of months within which to generate data.
That has been sent down to them from the top, and they have
to set up their machines at a certain speed. That speed is
not experimentally determined, but is determined from a
marketing perspective. What I would like to do is to focus
on one aspect of the setup to try and point out how
important it is that we focus our attentions on proper
experimental protocols, if we desire to generate valid
durability data. Of course, if your purpose is to just
generate data so that you can get a valve on the market,
well then if things happen down the road that you could have
predicted but didn't, because of your time constraints, that
is the kind of problem you are going to run into.
On the first slide you see a bioprosthetic reading. We
received a new valve a couple of weeks ago that we were
asked to generate protocols for, and I thought that this was
a particularly good example of what can go wrong if you are
not careful about what you do when you setup a valve. In
this particular instance, we have tested a 35 mm valve.
These are gigantic valves, and of course the larger the
valves are the more difficult they are to test, as you all
know. These are also a single occluder mechanical valves,
again very difficult by nature to tune.
What we have here is a valve that is being tested at 400
cycles per minute, pretty slow, most people would say.
Although it is difficult to see, we are running somewhere
between 118 and 125 mm Hg closing pressure. I took a series
of shots to show you that on every cycle the valve was
closing. We observed this microscopically, but you have to
really assess this with the pressure trace. This I expanded
just to show you that this is not an acoustic spike but a
true closing pressure. What we then did was try to speed up
the machine to 800 beats per minute, assuming that we were
not allowed to test this valve at 400. In actual fact, we
were not. At 800, we spent a couple of hours just trying to
tune the machine. The best that we could do by the criteria
I am talking about (no occluder flutter, closing on every
cycle, fairly reproducible closing pressure), is going to be
shown in these next four slides. You will find that the
closing pressures here are running anywhere from about 125
mm Hg of pressure to about 300 mm Hg of pressure. We are
still closing on every cycle, and most of our closing spikes
are running 125 to 175 mm Hg. We do have these transient
spikes that are coming up at about 250 mm Hg and sometimes
300 mm Hg. We had a big spike that is running at about 500
mm Hq. We found that about one out of every 100 cycles we
would reach about 500 mm Hg. The problem here, of course, is
that the occluders at this speed are simply not experiencing
reproducible motion. It is taking slightly more volume in one
cycle than the next to close these valves, which has to do
with the reproducibility of the motion of the occluder, and
the ability of the occluder to move with speed. We then
assumed that we were under the constraints of having to test
this valve at 1200 cycles per minute. Setting the machine at
1200 cycles per minute, we then adjusted all of the
parameters of the device to get the best traces that we
could. These are the best traces that we could get. On this
slide we have 125 mm Hg to about 300 mm Hg. On this
particular slide we get closing on every cycle. We took a
fast shot to show the nature of that spike. On this
particular one we are running anywhere from about 125 to 600
mm Hg closing pressure. We do get an average spike now and
then. The real trouble here is that on two of cycles the
valve has not closed enough to give us a closing pressure.
Now the point that I am trying to show here is that you have
to use your valve in your particular machine and go through
procedures of generating a valid protocol. You can't predict
ahead of time what speed you can test the valve at. I would
say of all of the testing laboratories that I have visited,
the biggest problem that I run into is the desire on the
part of the individuals running the machine to test the
valves faster than they are really able to do. It is not
just our instrument. It is all of the instruments that I
have observed. Again we run into the dilemma of wanting to
test valves appropriately, but we have time constraints put
upon us from the marketing end of the whole thing. What you
get into usually is a tendency to overlook certain valve
failures that you run into in an in-vitro manner. Because
you know that you have over tested the valve, you know that
you have some transient high pressure spikes, so that if 10%
of your spikes are higher than you want, and 10% of your
valves experience mechanical failure, you have a tendency to
overlook that and sure enough now we are working with
individuals who are losing millions of dollars a day because
they did that. All I can do is encourage you folks to
realize that if you want to generate valid accelerated
durability data, you really have to probably slow down your
testing. Thank you.
Discussion:
Q: ( Walker ) You mentioned cavitation earlier on, and my
question relates to that as it is a topic we have been
interested in for mechanical valves. I believe you said that
you try to eliminate it. In our experience we have seen
explanted valves where there has been evidence of
cavitation. I just wonder what your rationale is for
eliminating cavitation?
A: (Conti) We are also aware of recent information
indicating that cavitation can perhaps occur in-vivo, or at
least there is the ability to expose the valve in-vivo to
conditions in which material is removed from that valve in a
manner which makes it look as though it is experiencing
cavitation. It is rather shocking, I would think that you
would all agree to that. What I was referring to at the time
was a cavitation problem that was a little more wide spread
than the localized cavitation that we have all seen in our
explanted valves. It seems to us that if we are running
about 1200 cycles per minute and then slowing down to a
somewhat slower speed of about 1000, where we cannot observe
this wide spread banding of cavitation, that is probably
better for our testing. We have never observed, even on
valves that have experienced these problem in-vivo,
cavitation in the regions in which they do in-vivo.
Q: (Gabbay) Could you please explain the term cavitation.
A: (Conti) Cavitation has to do with pressure differential
that occurs in a flowing liquid which results in the actual
evaporation of the liquid adjacent to a surface. It is a
rapid expansion, constriction of an air bubble. (Here Conti
defers to Professor Swanson for a better explanation.)
A: (Swanson): A vapor is capable of performing - the damage
comes when it collapses - There is such a pounding when it
collapses, it will just eat out a molecule at a time. When a
small bubble collapse, as you see with heart valves, you see
this "flash" that occurs immediately. When you catch it with
a strobe, the flash is visible.
Q: (Gabbay) Why this air bubble does no real damage to
bioprostheses? Why only to mechanical valves?
A: (Swanson) Because the accelerations of the fluid locally
are not high enough in bioprostheses, as they are in the
mechanical valve. The mechanical valve closes and stops
suddenly, and you get very high pressure gradients of the
fluid. This is what causes it.
Q: (Gabbay) It could be that this cavitation is
characteristic of fatigue testers which have no compliance
or capacitance.
Q: (Wasserman) Which machine and which valve was that
tracing from?
A: (Conti) Well I can't tell you which valve that was from,
but it was from our six position durability tester. A
mechanical valve. A single occluder.
Q: (Wasserman) There is an abnormal amount of noise on these
curves. Did. you find it difficult to standardize from valve
to valve?
A: (Conti) Each valve is very unique in the problems that it
presents to the tester. I am sure you have experience that.
Mechanical valves always have more noise associated with
them. You are going to get some acoustic signals going
through the machine because of other valves closing. We do
have some electrical noise as well. Leaflet valves are
usually much easier to tune than the mechanicals, and so
their tracers do look somewhat different from that. In our
hands, every valve requires a special holder, and requires a
special protocol generated for it.