Experimental Considerations that Determine the Utility of Accelerated Durability Testing for Prosthetic Heart Valves
James C. Conti, Ph.D., Elaine R. Strope, Ph.D., Donald Rohde
Presented By: James C. Conti, Ph.D.
Dynatek Laboratories, Inc., Galena, MO
The reliability and usefulness of experimentally generated
data is dependent upon, among other things, the generation
of an experimental protocol that recognizes the importance
of pertinent variables. Any experiment that fails to
control or at least recognize those variables is probably
better off not being performed. This experimental sloppiness
is rampant throughout many of the world's prosthetic
heart valve testing laboratories. As a result of the uselessness
of the generated data an attitude has developed
that categorizes prosthetic heart valve and durability testing
as useless and irrelevant.
Experimental protocols can indeed be written that result in
the generation of durability data that is both relevant and
can be correlated with in vivo wear. This paper addresses
the steps that a researcher can follow to generate a meaningful
protocol.
The Testing Instrument
Regardless of the supplier or designer of the accelerated
tester, several assumptions are made in this paper.
1) Pressure measurements are free of inertial interference
and are
frequency independent over the range of
frequencies encountered.
2) The whole system has a near zero capacitance.
3) Pressure measurements do not change with time due
to instrument variation.
4) Pressure measurements reflect true load being
applied to valve.
General Considerations
There are three criteria that must be met for every valve
being tested. The first and second of these are that the
valve must fully open and then fully close each cycle.
This will of course require stroboscopic analysis. The
third criterion depends on the type of valve being tested.
A mechanical valve demands a lack of cavitation. A tissue
or polymer leaflet valve requires a lack of leaflet flutter.
If all the criteria cannot be attained it is usually
due to excessive testing speeds.
Leaflet Flutter
The most common flutter damage is fraying and tearing of
leaflets. Leaflets should smoothly close, appose, stop,
reverse and smoothly open. As a leaflet approaches the
fully opened position the chances for leaflet flutter are
maximized.
Cavitation
Cavitation occurs around the periphery of the occluder
during maximum motion and at the instant of closure. It
will appear as a momentary flash at the location of maximum
occluder motion. It can be reduced by lowering temperature,
flow rate, speed and raising systemic pressure.
Closing Pressures
The problem arises if the pressures measured are actually
higher than the load experienced by the valve at closure.
Researchers should therefore beware of situations in which
valve wear is less even though the same closing pressures
are measured.
Dynatek Laboratories, Inc., Fourth and Main, PO Box 254, Galena, MO 65656